Summary
The US Food and Drug Administration said it classified Boston Scientific’s correction of the OptiCross HD Coronary Imaging Catheter as a Class 2 recall. The company updated the instructions for use after reports of air embolism and of repeated flushing to restore image quality while the catheter was still in the patient.
Key Takeaways
- The FDA posted the recall on Nov 20, 2026, listed it as open, and identified use error as the cause of the recall.
- The recall covers all non-expired batches for six material numbers, with 1,288,662 total units in commerce and distribution across the US and internationally.
- Boston Scientific’s advisory directs users to clear air from the accessories first, then clear air from the catheter by completing all three flushing steps before insertion.
Boston Scientific Corporation initiated a correction for the OptiCross HD Coronary Imaging Catheter on Oct 22, 2026, to revise its instructions for use, according to the FDA’s recall notice. The FDA posted the Class 2 recall on Nov 20, 2026, and listed it as open under recall number Z-0510-2027 and recall event ID 100133.
The recall covers all non-expired batches for material numbers H74939352010, H74939352020, H74939352030, H74939352040, H74939352050, and H74939352060. The listed global trade item numbers are 08714729960744, 08714729960751, 08714729960768, 08714729960737, 00191506035419, and 00191506035426.
Boston Scientific updated the instructions in response to reports of air embolism and repeated flushing to restore image quality while the catheter was in the patient. The company identified air embolism as a risk, stating that intravascular air may cause air embolism with consequences of varying severity, including ST-segment changes, reduced blood flow, arrhythmia, myocardial infarction, cardiac arrest, and other serious injury. The FDA identified use error as the recall cause.
Boston Scientific sent customers an Important Medical Device Advisory dated Oct 22, 2026, directing them to review the air embolism updates in Appendix 1. The advisory directs users to complete device preparation and flushing before insertion, clearing air from all accessories first and then from the catheter by following all three flushing steps in the instructions for use. It tells users not to flush the catheter inside the patient if they suspect trapped air and to remove the catheter before flushing if additional flushing is required.
The company asked recipients to forward the notice to relevant personnel and facilities that received transferred products, and directed distributors to send it to their customers. The advisory also asks recipients to post the notice near affected products and return the acknowledgment form after completing it.
The recall lists 1,288,662 total units in commerce and worldwide distribution across the US and its territories, Europe, the Middle East, Africa, Canada, Japan, Latin America, and the Asia-Pacific region.